FLSC 05 w/FLI 225 Instrument Service Manual
59
9 Scale
Filter Type
1.
Select
Filter Type =
from the
Filter
module.
2.
Select from the available filtering options.
Filter2
is the default setting.
None
means no filtering.
Filter1
filtering is slower response to weight in a longer time period with
improved accuracy.
Filter2
filtering is faster response to weight in a short time.
3.
Press the
Enter
softkey to accept and return to the
Filter
screen.
Constant
If you turn filtering on, you need to set the
Constant
. Typical values are between 1 and
10 (default = 4). Set the number low for small vibration problems and higher for more
dampening effect.
1.
Select
Constant
from the
Filter Module
.
2.
Press the
Up
or
Down Arrow
softkey to choose a number between 0 and 10.
3.
Press the
Enter
softkey.
The constant is stored and the instrument returns to the
Filter
screen.
Threshold
The purpose of the Threshold is so the instrument will respond quickly to large weight
changes. Threshold is the amount of weight change, in calibration units, beyond which
the filter will be temporarily disabled. For example, if you set this to 10 lbs, a weight
change over 10 pounds occurring during the sample time (½ sec. in our example) will
disable the filter until the weight change during the sample time drops below 10 lbs.
1.
Select
Threshold
from the
Filter Module
.
2.
Scroll in the value in calibration units using the data entry procedure
(Default = 3).
3.
Press the
Enter
softkey to return to the
Filter
screen.
9.2.4 Motion/AZT
1.
Select
Motion/AZT
from the
Scale
module to display the following screens.
Figure 9.32 Motion/AZT Screens
2.
After defining the settings, press the
Esc
softkey to exit and return to
Scale
.
Summary of Contents for FLI 225
Page 10: ...10 FLSC 05 w FLI 225 Instrument Service Manual ...
Page 11: ...FLSC 05 w FLI 225 Instrument Service Manual 11 1 FCC and EMC Declarations of compliance ...
Page 70: ...70 FLSC 05 w FLI 225 Instrument Service Manual Chart of ASCII Characters ...
Page 106: ...106 FLSC 05 w FLI 225 Instrument Service Manual 11 5 Exploded View of FLI 225 AWT05 500348 ...
Page 109: ......