76
FLSC 05 w/FLI 225 Instrument Service Manual
9.7.3 Passwords
1.
Select
Passwords
from the
Miscellaneous
module to access the
Passwords
screen.
Figure 9.48 Passwords Screen
2.
Select the password you wish to change from the list.
The current password for that option is displayed.
3.
Scroll in the new password using the
Data Entry Procedure on page 13
.
4.
Press the
Enter
softkey to accept the new password and return to the
Passwords
module.
9.7.4 Display
1.
Select
Display
from the
Miscellaneous
module to access the
Display
screen.
Figure 9.49 Display Screen
2.
Select
Update Rate (Hz) =
to access the list of available update rates.
3.
Change the update rate to
1
,
2
,
5
or
10
Hz as necessary. The default is
5
.
The change is made and the instrument returns to the
Display
screen.
4.
Select
Separator =
from the
Display
screen to display the available options
(decimal point or comma).
5.
Select the decimal separator as necessary.
The change is made and the instrument returns to the
Display
screen.
6.
Press the
Esc
softkey to return to the
Miscellaneous
module.
CAUTION: Changing the default Admin password is not recommended as you
will not be able to access the menus if it is lost or forgotten.
Summary of Contents for FLI 225
Page 10: ...10 FLSC 05 w FLI 225 Instrument Service Manual ...
Page 11: ...FLSC 05 w FLI 225 Instrument Service Manual 11 1 FCC and EMC Declarations of compliance ...
Page 70: ...70 FLSC 05 w FLI 225 Instrument Service Manual Chart of ASCII Characters ...
Page 106: ...106 FLSC 05 w FLI 225 Instrument Service Manual 11 5 Exploded View of FLI 225 AWT05 500348 ...
Page 109: ......