FLSC 05 w/FLI 225 Instrument Service Manual
35
9
Change Seal Status
The FLI 225 has a Category 3 audit trail. For locations requiring a physical seal, you
must unseal the instrument before accessing the
Admin Management
menu.
1.
Access the
User Settings
module using password 111.
2.
Press the
Up
or
Down Arrow
softkey as necessary to highlight
Seal Status
.
3.
Press the
Enter
softkey.
The current status (sealed or unsealed) is displayed.
Figure 9.2 Sample Seal Status Screen
4.
If the status is unsealed, skip to step 8. Otherwise continue with the following.
5.
Remove the back panel of the instrument.
6.
Press the seal switch located at S1 on the communications/power supply
board.
The status changes from sealed to unsealed.
7.
Replace the back panel.
8.
Press the
Esc
or
Enter
softkey to return to the previous module.
The display returns to the
User Settings
screen.
9.
Press the
Esc
softkey 2 times to exit to normal operating mode.
10.
Access the
Setup Manager
module as explained on page 34.
11.
After completing all changes in the
Setup Manager
, reseal the instrument as
necessary. The steps for sealing the instrument are the same as for unsealing.
Summary of Contents for FLI 225
Page 10: ...10 FLSC 05 w FLI 225 Instrument Service Manual ...
Page 11: ...FLSC 05 w FLI 225 Instrument Service Manual 11 1 FCC and EMC Declarations of compliance ...
Page 70: ...70 FLSC 05 w FLI 225 Instrument Service Manual Chart of ASCII Characters ...
Page 106: ...106 FLSC 05 w FLI 225 Instrument Service Manual 11 5 Exploded View of FLI 225 AWT05 500348 ...
Page 109: ......