FLSC 05 w/FLI 225 Instrument Service Manual
39
9 Calibrate the FLSC 05 with FLI 225
7.
Place 500 pounds on each of the corners in succession, noting the weight
reading for each corner.
Figure 9.5 Weight on corners
If the system fails this test, perform the Corner calibration as explained
on page 48 using a minimum of 500 pounds for the span weight before
continuing with step 8.
8.
Place 500 pounds in the center of the pallet.
9.
Tilt the mast forward and note the weight reading.
10.
Tilt the mast back and note the weight reading.
If the system fails either step 9 or step10, perform the 16 point Angle
Calibration as explained on page 51 using a minimum of 2000 pounds
for the span weight.
11.
Upon completion, reconfigure the indicator to 5 lb divisions and AZT - ON.