30
FLSC 05 w/FLI 225 Instrument Service Manual
2.
Press the
Up
or
Down Arrow
softkey as necessary to highlight the entry you
wish to see.
3.
Press the
Enter
softkey.
The details for that entry are displayed.
Figure 7.10 Sample Error Log Record
4.
Press the
Esc
or
Enter
softkey
to return to the
Error Log Viewer
.
5.
Press the
Esc
softkey to the
Diagnostics Manager
module.
The following errors may be logged if enabled.
Table 7.1 Error Codes and Definitions
Error
Definition
UNDERLOAD ERROR The weight on the carriage has gone below the set negative% of capacity.
OVERLOAD ERROR The weight on the carriage has gone above the set positive% of capacity.
DRIFT ERROR
Zero Tracking has removed an accumulated weight beyond the set% (positive or negative) of
capacity.
FLJ COMM ERROR Communication with the FLJ 100 or FLW 100 has been lost.
FLJ CHANGE ERROR An FLJ 100, FLW 100 or FLI 225 instrument has been changed.
FLJ CORNER ERROR Stored corner coefficients do not match between FLI 225 and FLJ 100 or FLW 100
FLJ ANGLE ERROR Stored angle coefficients do not match between FLI 225 and FLJ 100 or FLW 100.
Summary of Contents for FLI 225
Page 10: ...10 FLSC 05 w FLI 225 Instrument Service Manual ...
Page 11: ...FLSC 05 w FLI 225 Instrument Service Manual 11 1 FCC and EMC Declarations of compliance ...
Page 70: ...70 FLSC 05 w FLI 225 Instrument Service Manual Chart of ASCII Characters ...
Page 106: ...106 FLSC 05 w FLI 225 Instrument Service Manual 11 5 Exploded View of FLI 225 AWT05 500348 ...