42
FLSC 05 w/FLI 225 Instrument Service Manual
After Replacing the FLI 225
Calibration information is stored in both the junction box (FLJ 100) and the instrument.
If the FLI 225 is replaced, you can retrieve the stored calibration information from the
junction box.
1.
After connecting the FLI 225 to the system, power the instrument on.
The following screen is displayed.
Figure 9.6 FLJ 100 Coefficient Mismatch Screen 1
2.
Press the
No
softkey.
The second option is displayed.
Figure 9.7 FLJ 100 Coefficient Mismatch Screen 2
3.
Press the
Yes
softkey to retrieve the stored calibration coefficient information
from the junction box.
The parameters are saved and the instrument returns to normal
operating mode.
4.
Press the
ZERO
key if the weight displayed is not zero.
5.
Proceed to one of the calibration tests on pages 38 through 41.
If the screen in Figure 9.6 does not display on power up, the FLI 225 that is connected
to the system was set to Factory Defaults. Calibration information will automatically
be uploaded from the junction box. Proceed to one of the calibration tests.
For instruments with EU configuration, selecting No for both options will lock the
instrument to prevent weighing until calibration has been completed. The instrument
will display dashes and display the message, “Calibration Required.”
Summary of Contents for FLI 225
Page 10: ...10 FLSC 05 w FLI 225 Instrument Service Manual ...
Page 11: ...FLSC 05 w FLI 225 Instrument Service Manual 11 1 FCC and EMC Declarations of compliance ...
Page 70: ...70 FLSC 05 w FLI 225 Instrument Service Manual Chart of ASCII Characters ...
Page 106: ...106 FLSC 05 w FLI 225 Instrument Service Manual 11 5 Exploded View of FLI 225 AWT05 500348 ...
Page 109: ......