FLSC 05 w/FLI 225 Instrument Service Manual
49
9 Calibrate the FLSC 05 with FLI 225
3.
Place weight on the rear, right corner of the pallet.
Figure 9.21 Corner Calibration Point 1
The instrument displays both the live and the most recent stored raw
count values for the cells.
Figure 9.22 Sample Live and Stored Counts
4.
Wait for counts to stabilize
5.
Press the
Store
softkey.
The instrument stores the counts and returns to the
Set/View Corner
Points
screen.
6.
Remove the weight (except for the deadload).
Pt 2: Left Rear - Pt 4: Right Front
The instructions for points 2 through 4 are the same as those for point 1 and are
described in this section. The instructions continue from the previous section.
7.
From the
Set/View Corner Points
screen, press the
Down Arrow
softkey to
highlight the next point (
Pt 2: Left Rear, Pt
3:
Left Front
, etc.)
8.
Press the
Enter
softkey.
The instrument displays the live and stored raw count values for the
cells.
The last two or three digits of the “Live” counts value will continue to fluctuate.