100
FLSC 05 w/FLI 225 Instrument Service Manual
10.8 SMA Protocol
Table 10.3 SMA protocol
Command Sent to Indicator
Result
<LF>W<CR>
Weight returned if no motion. Dashes displayed if motion on scale.
<LF>P<CR>
Weight returned if no motion. Dashes displayed if motion on scale.
<LF>Z<CR>
Scale zeros itself if no motion. Dashes displayed if motion on scale.
<LF>T<CR>
Scale tares itself and returns G or N weight if no motion. Dashes displayed if motion on scale.
<LF>T<xxxxxx.xxx><CR>
Scale attempts to take the <xxxxxx.xxx> data as the tare weight and returns G or N weight if no
motion. Dashes displayed if motion on scale.
<LF>M<CR>
Returns the tare weight if no motion. Dashes displayed if motion on scale.
<LF>C<CR>
Clears the tare weight if no motion. Dashes displayed if motion on scale.
<LF>U<uuu><CR>
Sets the unit of measure label to uuu Example: lb_ ( _ =space) Works for lb or kg only, not
custom units
<LF>D<CR>
Runs scale diagnostics and sends diagnostic message
<LF>A<CR>
Sends the SMA compliance level. See note at left.
<LF>B<CR>
1st B sent returns manufacturer
2nd B sent returns model software #
3rd B sent returns the software revision level
4th B sends an END
5th or more sends a ?
<LF>U<CR>
Toggles the units of measure if no motion. Dashes displayed if motion on scale.
<LF>I<CR>
Sends the SMA compliance level as
SMA:compliance level / revision level
Summary of Contents for FLI 225
Page 10: ...10 FLSC 05 w FLI 225 Instrument Service Manual ...
Page 11: ...FLSC 05 w FLI 225 Instrument Service Manual 11 1 FCC and EMC Declarations of compliance ...
Page 70: ...70 FLSC 05 w FLI 225 Instrument Service Manual Chart of ASCII Characters ...
Page 106: ...106 FLSC 05 w FLI 225 Instrument Service Manual 11 5 Exploded View of FLI 225 AWT05 500348 ...
Page 109: ......