OPERATION
30
RETURN CABLE CONNECTION
For the return cable connection, it is suggested that two insulated banana plugs be used
with 18AWG stranded wire. For proper connection, one end of the return cable should be
connected to the RETURN jack on the front panel of the Hypot II and the other end
connected to the INTERCONNECT jack on the front panel of the HYAMP jr. The
suggested E.F. Johnson part number for the banana plug is as follows:
108-0303-001
Banana Plug Insulated (2)
2. INTERCONNECTING WITH THE HAJA-01 ACCESSORY CABLE
Accessory kit HAJA-01 includes the rear interface cables needed for interfacing the
HYAMP jr. with the Hypot II . The following table lists the kit contents.
Part Number
Description
Qty.
HS-8-11
Cable Assembly Return Interconnect
1
3030D-22
Cable Assembly R/P Interface Sequential
1
3030D-23
Cable Assembly R/P Interface Simultaneous
1
37893
Adapter Box High Voltage High Current
1
The following sections describe the procedures for interconnecting your units. Please
refer to the test method you require for specific cable connection instructions.
Cable Connections.
First locate the HS-8-11 return cable, identified by its double banana plug connectors.
Connect one end to the RETURN jack on the front panel of the Hypot II and the other end
to the INTERCONNECT jack on the front panel of the HYAMP jr.
Determine whether you want to run the Ground Bond and Hipot test simultaneously or
sequentially then select the appropriate cable assembly for your needs, using the
descriptions in the table above. Attach the male connector with two cables coming from
it, to the female output connector of the HYAMP jr. Attach the shorter of the two
remaining cables with female connectors, to the male input connector of the HYAMP jr.
and the longer cable to the male signal input/output connector on the rear panel of the
Hypot II.
RETURN
INTERCONNECT
(
CHASSIS)
INPUT/OUTPUT
Hypot II
HYAMP jr.
REAR PANEL
REAR PANEL
FRONT PANEL
FRONT PANEL
HV
RTN
HC
Device Under Test
N
L
G
OUTPUT INPUT
Basic Test Setup
Summary of Contents for 3030D
Page 4: ......
Page 5: ...SECTION 1 OPERATORS MANUAL ...
Page 38: ...OPTIONS 34 This page left intentionally blank ...
Page 39: ...SECTION 2 SERVICE MANUAL ...