OPERATION
23
milliohms with a range from 0-510m
Ω
.
3.
Dwell Time Setting
Please press the SET key until arriving at the Timer setting. The display will show:
T i m e r = XXX . X s
Use the Up (
∧
) or Down (
∨
) arrow keys to adjust the Timer setting, then press the EXIT
key to store the Timer setting and return to the Run mode, or press the SET key to
advance to the next parameter. The range is 0 to 999.9 seconds, if the Timer is set to 0,
the instrument will operate in a continuous ON mode when the test button is depressed
and released. It will stop when the DUT (Device Under Test) goes into failure or the
manual reset button is pressed.
4. Frequency Setting
Please press the SET key until arriving at the Frequency setting. The display will show:
Freq = 50 HZ
or
Freq = 60 HZ
Use the Up (
∧
) or Down (
∨
) arrow keys to change the Frequency setting to 50Hz or 60Hz,
then press the EXIT key to store the frequency setting and return to the Run mode, or
press the SET key to advance to the next parameter.
5. Offset Setting
Please press the SET key until arriving at the Offset setting. The display will show:
Offset = XXX m
Ω
TEST to Auto Set
Please use the Up (
∧
) or Down (
∨
) arrow keys to adjust the setting of the offset manually,
then press the EXIT key to store the setting and return to the Run mode, or press the SET
key to advance to the next parameter. To automatically set the Offset, press the TEST
button when the test leads or test fixture is connected across the Hyamp jr. outputs. The
HYAMP jr. will measure the resistance that is present and enter this value into the Offset
setting. This offset will be automatically subtracted from the actual resistance reading
before the reading is displayed. The unit of measure is in milliohms with a range from 0-
100m
Ω
.