CONTROLS
14
REAR PANEL CONTROLS
1
2
3
4
5
6 7
8
1.
THERMAL FAN:
Runs continuously to cool the instrument.
2.
INPUT POWER SWITCH:
Line voltage selection is set by the position of the
switch. In the left position it is set for 115-volt operation, in the right position it is set
for 230-volt operation.
3.
REMOTE OUTPUT:
9 pin D subminiature female connector for monitoring PASS,
FAIL, and PROCESSING output relay signals, as well as remote START and RESET
outputs.
4.
CALIBRATION ENABLE:
This recessed switch when depressed allows access to
the calibration control program.
5.
REMOTE INPUT:
9 pin D subminiature male connector for remote control of test
and reset functions as well as withstand processing input.
6.
CHASSIS GROUND (EARTH) TERMINAL:
This safety terminal should be
connected to a good earth ground before operation.
7.
FUSE RECEPTACLE:
To change the fuse unplug the power (mains) cord and
unscrew the fuse cap counter clockwise to remove the fuse.
8.
INPUT POWER RECEPTACLE:
Standard IEC 320 connector for connection to a
standard NEMA style line power (mains) cord.