QUICK START
18
QUICK START
This quick start guide assumes the operator has some familiarity with
automated Ground Bond testing and desires to use the
default
settings
on the instrument. The default settings shown will remain in memory
unless you choose to override them with your own test program. The
instrument default settings are as follows:
DEFAULTS
•
Memory Position:
1
•
Input
Voltage:
115 volts AC (rear panel switch selectable)
•
Current
Output:
25.00 Amps
•
Resistance Trip (High):
100m
Ω
•
Dwell
Timer:
1 second
•
Frequency:
60Hz
•
Offset:
20m
Ω
•
Lock:
Key Unlock
a). Unpack this instrument from its special shipping container. Be sure to save all
packaging materials in case you need to return it to the factory for service.
b). Locate a suitable testing area and be sure you have read all safety
instructions for the operation of the instrument and suggestions on the
test area set-up in the SAFETY section of this manual. Locate a three prong grounded
outlet. Be sure the outlet has been tested for proper wiring before connecting the
instrument to it.
c). Check to be sure that the correct input line voltage has been selected
on the rear panel, either 115 volts AC or 230 volts AC. Connect the
power-input plug into its socket on the rear panel of the instrument. Connect the male
end of the plug to the outlet receptacle.