QUICK START
20
to quickly ABORT a test and cut off the HIGH CURRENT.
When HIGH CURRENT is present a green TEST indicator located to the left of the
CURRENT receptacle will light until the HIGH CURRENT is shut OFF. If the device
under test PASSED, then a single beep will sound. In the case of a FAIL condition, the
RESET SWITCH will illuminate and the LCD will display a failure
condition along with an audible alarm. The instrument will provide a
memory of the test condition results on the LCD display that will remain
until the next test is initiated. Depressing the reset button will reset the instrument alarm
while keeping the last test results on the display. Depressing the reset button a second
time will clear the display.
Quick Guide to Interconnecting HYAMP jr. to Hypot II
Using the Supplied Interface Cable Kit HAJA-01.
1.
Connect the double ended banana plug cable from the Hypot II front RETURN jack to
the HYAMP jr. INTERCONNECT jack. Determine whether you want to run the
Ground Bond and Hipot test simultaneously or sequentially then select the appropriate
cable assembly for your needs, using the descriptions in the table below.
Cable number
Method
Operation
3030D-22
Sequential
HYAMP jr. first then Hypot II.
3030D-23
Simultaneous
HYAMP jr. and Hypot II at the same time.
Note: The cables are identified on the cable label.
2.
Attach the male connector with two cables coming from it, to the female output
connector of the HYAMP jr.. Attach the shorter of the two remaining cables with
female connectors, to the male input connector of the HYAMP jr. and the longer
cable to the male signal input/output connector on the rear panel of the Hypot II
3.
Connect the spade lug of the high current lead (red) from the (optional) receptacle
adapter box (37893) to the HYAMP jr. jack marked CURRENT (red), and the high
voltage plug (white) to the Hypot II jack marked HV (white).
4.
Connect the RETURN lead with the alligator clip (black)
to the HYAMP jr. jack
marked RETURN (black). The alligator clip should be connected to the DUT chassis.
5.
Check the test parameters and initiate the tests by pressing the HYAMP jr. test button.
The Reset buttons on both instruments are active to disable the tests or for failure
reset.
•
For a more detailed description on interconnecting your system, see page 28.