SETTING FREQUENCY AND AMPLITUDE
b. For better voltage and dBm resolution, set Meter Function
(1)
to LEVEL. The meter
(3)
is read in conjunction
with the OUTPUT LEVEL controls
(10)
(e.g., with OUTPUT LEVEL set to approximately 20 mVOLTS, a meter
reading of 2.1 indicates that the actual level is 21 mVrms).
c. If a 50 ohm to 75 ohm adapter (consisting of a 25 ohm series resistor) is connected to RF OUTPUT
(9),
the
OUTPUT LEVEL
(10)
voltage scale will be correct if the instrument is used with 75 ohm terminations. However,
1.76 dB must be subtracted from the dB scale for correct readings.
Figure 3-5. Setting the Frequency and Amplitude Controls (3 of 3)