Model 8640B
TM 9-4935-601-14-7&P
Table 4-1. Performance Test Record (6 of 8)
Results
Para
Test Description
No.
Min
Actual
Max
4-35
Peak Incidental Phase Modulation Test
512 MHz
______
+17.20
128 MHz
______
+8.60
4-36
Pulse Modulation Test
0.5-1 MHz
Rise Time
______
9 µs
Fall Time
______
9 µs
Level Accuracy
5.4 div
______
6.7 div
1-2 MHz Rise Time
______
4 µs
Fall Time
______
4 µs
Level Accuracy
5.4 div
______
6.7 div
2-4 MHz Rise Time
______
2 µs
Fall Time
______
2 µs
Level Accuracy
5.4 div
______
6.7 div
4-8 MHz Rise Time
______
2 µs
Fall Time
______
2 µs
Level Accuracy
5.4 div
______
6.7 div
8-16 MHz Rise Time
______
1 µs
Fall Time
______
1 µs
Level Accuracy
5.4 div
______
6.7 div
16-32 MHz
Rise Time
______
1 µs
Fall Time
______
1 µs
Level Accuracy
5.4 div
______
6.7 div
32-64 MHz
Rise Time
______
1 µs
Fall Time
______
I1 s
Level Accuracy
5.4 div
______
6.7 div
64-128 MHz
Rise Time
______
1 µs
Fall Time
______
1 ,µs
Level Accuracy
5.4 div
______
6.7 div
128-256 MHz Rise Time
______
1 µs
Fall Time
______
1 µs
Level Accuracy
5.4 div
______
6.7 div
256-512 MHz
Rise Time
______
1 µs
Fall Time
______
1 µs
Level Accuracy
5.4 div
______
6.7 div
4-37
Pulse ON/OFF Ratio Test
Frequency Range: 256-512 MHz
40 dB
______
128-256 MHz
40 dB
______
64-128 MHz
40 dB
______
32-64 MHz
40 dB
______
16-32 MHz
40 dB
______
8-16 MHz
40 dB
______
4-8 MHz
40 dB
______
2-4 MHz
40 dB
______
1-2 MHz
40 dB
______
0.5-1 MHz
40 dB
______
4-82
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com