Model 8640B
TM 9-4935-601-14-7&P
ADJUSTMENTS
5-37. RF FILTER ADJUSTMENT (Cont'd)
Figure 5-5. RF Filter Adjustment Test Setup
EQUIPMENT:
Spectrum Analyzer........................................................... HP 141T/8552B/8554B
Tracking Generator .............................................................................. HP 8444A
Frequency Counter...............................................................................HP 5327C
Test Cable (2 required) ..............................................................HP 11592-60001
Adapter ......................................................................................... HP 1250-0827
PROCEDURE:
1. Connect equipment as shown in Figure 5-5 after setting Signal Generator's controls as follows:
RANGE ......................................................................................... 256 - 512 MHz
FREQUENCY TUNE. ...................................................... ........................Fully cw
RF ON/OFF ..................................................................................................OFF
2. Set spectrum analyzer center frequency to 550 MHz, frequency span (scan width) to 100 MHz per division,
resolution bandwidth to 10 kHz, and input attenuation to 20 dB.
3. Set tracking generator's output level to 0 dBm. Adjust the tracking for maximum response in a 10 kHz resolution
bandwidth. (Tracking should be checked periodically during this test.) Set analyzer's resolution bandwidth to 300
kHz.