Model 8640B
TM 9-4935-601-14-7&P
Table 1-2. Recommended Test Equipment (7 of 7)
Instrument Type
Critical Specifications
Suggested Model
Use*
Tracking Generator
Output: to 0 dBm (50
Ω
)
HP 8444A
P,A
Flatness: ±0.5 dB
Compatible with Spectrum Analyzer
(HP 141T/8552B/8554B)
Tracking Generator
Output: to 0 dBm (50
Ω
)
HP 8443B
P,A
Compatible with Spectrum Analyzer,
(HP 141T/8552B/8553B)
Variable Phase Oscillator
Range: 20 Hz to 60 kHz
HP 203A
P,A
Output Impedance: 600
Ω
Phase Variability: 0 to 360°
Distortion: >64 dB down
Variable Voltage
Range: +5% to -10% of nominal
GR W5MT3A
P
Transformer
line voltage (100, 120, 220 or
240 volts). For 120V, range
is 105 - 130 Vrms.
Metered Accuracy: ±1 Vrms
Vector Voltmeter
Range: 1 - 550 MHz
HP 8405A
P
Sensitivity: <20µV
Phase Range: +18° full scale down
to +6° full scale
Phase Resolution: 0.1°
Phase Accuracy: ±1.5°
Voltage Ratio Accuracy: 0.2 dB
VSWR Bridge
Range: 0.45 - 550 MHz
Wiltron Model
P
Directivity: >40 dB
60N50
Connectors: Type N
*P = Performance; A = Adjustments; T = Troubleshooting
1-19