TM 9-4935-601-14-7&P
PERFORMANCE TESTS
4-29. OUTPUT LEAKAGE TEST (Cont'd)
5. Set the analyzer's center frequency controls to 700, 900, and 1100 MHz. Hold the end of the loop antenna
cylinder in contact with all surfaces of the generator at each frequency setting. All signals and noise should be
below the -27 dB graticule line (below -97 dBm) from 600 MHz to 1200 MHz.
-97 dBm____________
4-30. INTERNAL MODULATION OSCILLATOR TEST
SPECIFICATION:
Standard:
Frequency: fixed 400 Hz and 1000 Hz ±2%.
Output Level: indicated 10 mVrms to 1 Vrms into 600 ohms.
Option 001:
Frequency: variable 20 Hz to 600 kHz +10% in 5, decade continuous bands plus fixed 400 Hz and 1000 Hz ±2%.
Output Level: 20 mVrms to 3 Vrms into 600 ohms.
DESCRIPTION:
The internal modulation oscillator output is measured with a voltmeter and a frequency counter to verify its frequency
range and accuracy and its level.
Figure 4-16. Internal Modulation Oscillator Test Setup.
EQUIPMENT:
Frequency Counter...............................................................................HP 5327C
Digital Voltmeter ....................................................................... HP 3480B/3484A
600 Ohm Feedthrough Termination ................................................... HP 11095A
4-44