6.7
OPERA
TION
[1]
The Plant Zone (PZ) is the space, within the Upper
Compartment of the Sampler Cabinet.
[2] The Container Zone (CZ) is the space, within the Lower
Compartment of the Sampler Cabinet.
Taking a Spot Sample
With the Display showing the
DEFAULT DISPLAY
, Press the
SPOT
SAMPLE
button on the Keypad. The Sampler should extract a Single
Shot Sample of water from the Sampling Point and deposit it in the
Sample Collection Vessel, according to the
SAMPLING CYCLE
.
The
SAMPLING
CYCLES
for the basic Aquacell Sampler and the
Aquacell Sampler in PPI mode are as follows:
Switching on the (Optional) Back-Up Battery
Basic
(For S50, S100 and S200 Samplers)
1. Release the Latch at the rear of the Sampler Module and open the
hinged Battery Cover.
2. Connect the Battery to it's cable using the red and black
connector.
S300 Series
1. Switch the Back-Up Battery Switch on the Cabinet Front Panel,
above the Sampler Module to the CONNECTED position.
CAUTION
Always disconnect the Back-Up Battery during prolonged (> 24 hours)
power outage, or damage could result.
S310H
S320
S320H
Plant Item
Location in the
Sampler Cabinet
Action At
Switch-on
Fridge
In the Plant Zone
[1]
Starts to
run after an
initial delay
of 3 minutes,
provided
the ambient
temperature is
above 3˚C
Summary of Contents for Aquacell S100