6.40
OPERA
TION
Leaving the Sampler to Operate Automatically
•
Ensure that the Sample Collection Vessel(s) is (are) clean and dry.
•
Locate the Sample Collection Vessel taking care to ensure that
the Samplers Pinch Valve Tube (the Sample Discharge Point) is
properly located in the Collection Vessel point of entry.
•
Switch on the power.
•
Program the Sampler as required.
•
Take a few Spot Samples to verify operation (you may wish to use
a substitute Sample Collection Vessel to collect these samples in
order not to contaminate the Sample Collection Vessel(s) to be
used for the sample program).
•
Press the
RUN PROG
button.
•
Close and secure the Front Cover / Door as appropriate.
•
Leave the Sampler to extract its Programmed Samples.
•
NB You may find the Test Mode (page 7.5) useful when setting up
the Sampler).
Returning to Collect Samples
• Open the Front Cover / Door.
• Press the
VIEW PROGRAM
Button to check the Sampler’s
movements.
• Press the
STOP
Button.
• Remove the Sample Collection Vessel(s).
• Either decant the Samples into separate containers or retrieve the
charged Sample Collection Vessel(s) intact as required. NOTE
(Samplers incorporating Sample Temperature Control) Provision
should be made to maintain Samples at a nominal 4°C during
transportation back to the laboratory.