4.3
PREP
ARING
THE
SITE
(Optional) Ancillary Signal Connection
See Connecting the Sampler to Typical Auxiliary Equipment, Aquacell
S50 Sampler (page 4.4). Auxiliary Equipment should, if possible, be
positioned adjacent to where it is intended to mount the Sampler,
within reach of the Samplers Ancillary Signals Cable.
Cables to / from the Auxiliary Equipment may need to be terminated
in one or more Signal Isolators (Not supplied by Aquamatic) mounted
adjacent to the Sampler. The isolators should be connected to the
Sampler via the Samplers Ancillary Signals Cable.
Isolators are necessary to protect the Samplers inputs / outputs from
damaging electrical transients, which are sometimes induced in long
cables.