4.14
PREP
ARING
THE
SITE
Sampler Location
A flat, horizontal mounting base (typically a 100mm thick concrete
pad) should be provided for the Sampler, preferably with a standing
area in front and service access to the rear.
Gravity Drain (Condensate) Installation (if required)
Construct the drain from 21.5mm O/D plastic pipe.
It is important when configuring the drain to avoid the possibility of
water being removed from the trap by siphon action.
The Sampler is provided with a 21.5mm O/D stub pipe at the bottom
rear of the Sampler Cabinet. The drain-pipe is joined to the stub pipe
using a suitable fitting.
(Optional) Condensate Evaporator Tray Installation
Attach the 4 Thermal Isolation Springs to the feet of the tray. The
springs are quite a tight force fit. The tray gets too hot to touch and
the Thermal Isolation Springs serve to isolate the hot under-surface of
the tray from the floor that it stands on.
The tray stands on the Mounting Base at the rear of the Sampler
suitably positioned beneath the Condensate Drain Outlet Stub Pipe.
In this way condensate is deposited into the tray and evaporated
off into the atmosphere. The Condensate Evaporator Tray is only
energised when the Sampler’s Refrigeration Unit is operating (to
minimise electricity consumption).
Gravity Drain (Wastewater) Installation (if required)
Construct the drain from 56mm O/D plastic pipe and terminate it in
the P-Trap supplied with the Sampler (attach the drainpipe to the
output of the P-Trap i.e. the lower connection point).
It is important when configuring the drain to avoid the possibility of
water being removed from the trap by siphon action.
The Sampler is provided with a 56mm stub pipe at the bottom rear of
the Sampler Cabinet. The P-Trap is attached to the Sampler at this
point.