Fuses protect the critical electrical devices on Aqua Phase from electrical anomalies.
Without a
properly fused electrical system, there would be risks to the critical (expensive) electrical
devices and to
the
person operating the device.
IF YOU SUSPECT A FUSE MAY BE BLOWN, UNPLUG THE MACHINE BEFORE
INVESTIGATING.
Seven fuses and a GFCI are located on
Aqua Phase:
Name
Fuse Type
Purpose
Location
FU-1
RFK 20A
(FRN-R-20) Pump Motor - Main Power
Control Panel
FU-2
CC 5A
(FNQ-R-5)
Electronics System
Control Panel
FU-3
GMD 2.5A
(GMD-2.5A) PLC Outputs
Control Panel
FU-4
GMD 1.5A
(GMD-1.5A)
Optional
Drive Motor
Control Panel
FU-5
GMD 1.5A
(GMD-1.5A) Safety Switches & Probes
Control Panel
FU-6
GMD 1.5A
(GMD-1.5A) PLC Inputs (Switches)
Control Panel
FU-7
CC 8A
(FNQ-R-8)
Drain Pump
Control Panel
FU-8
-
Optional (Lite, Blower, etc) Control Panel
GFCI
GFCI
Ground Fault Protector
Power Cord
Effects of a blown fuse(s):
Name
Status
Condition
FU-1
If Blown
No power to Main Pump, no water spraying in Wash Chamber.
FU-2
If Blown
System will be dead.
FU-3
If Blown
No activity from pressing “
START
” button.
FU-4
If Blown
System will pump, but no sweeping spray action will be heard.
FU-5
If Blown
No activity from pressing “
START
” button.
FU-6
If Blown
No activity from pressing
"START"
button.
FU-7
If Blown
No Power to drain pump. Machine will not drain.
GFCI
If Tripped
System will be dead. No power to any devices.
Fuses listed above with the exception of the “GFCI” are located behind the control panel face.
To access fuses on the control panel:
1.
Remove the fasteners on the left and right of the control panel cover.
2.
You need to
remove only the top two
screws on the control panel face.
FUSES