38
□
To tare the weighed value when weighed value is stable (no response if
erCd 0
)
Command T C
R
L
F
Response
<AK>
C
R
L
F
When zeroing operation is possible
□
To switch the weighing unit (no response if
erCd 0
)
Command U C
R
L
F
Response
<AK>
C
R
L
F
When zeroing operation is possible
□
To set the ID number (no response if
erCd 0
)
Command I D : 0 0 0 1 0 0 C
R
L
F
Response
<AK>
C
R
L
F
□
To request the ID number
Command ? I D C
R
L
F
Response I D , 0 0 0 1 0 0 C
R
L
F
□
To set the HI limit value (no response if
erCd 0
)
Sends a six-digit number that does not contain “+/-” or decimal point
Command H I : + 0 0 0 5 0 0 C
R
L
F
Response
<AK>
C
R
L
F
□
To set the LO limit value (no response if
erCd 0
)
Sends a six-digit number that does not contain “+/-” or decimal point
Command L O : + 0 0 0 2 0 0 C
R
L
F
Response
<AK>
C
R
L
F
□
To set the comparator memory (no response if
erCd 0
)
Command C N : X X C
R
L
F
XX is memory number (00
~
09
)
Response
<AK>
C
R
L
F
□
To output the HI limit value
Command ? H I C
R
L
F
Response H I , + 0 0 0 5 0 0 C
R
L
F
□
To output the LO limit value
Command ? L O C
R
L
F
Response L O , + 0 0 0 2 0 0 C
R
L
F
□
To recall the comparator memory
Command ? C N
Response C N , X X C
R
L
F
XX is memory number (00
~
09
)
□
If the command cannot be executed or the received command cannot be processed, the
following error code will be returned.