25
14.4.
Restoring Sensitivity Adjustment Value to Factory Default
To change sensitivity adjustment due to incorrect operation, etc., restore sensitivity adjustment by
gravity acceleration or weight to factory default.
1. Refer to “
14.1. Sensitivity Adjustment Items
Cal
is
displayed when you enter sensitivity adjustment mode.
2. Press the MODE key two times to display
ClrC
.
3. Press the TARE key to display
ClrC no
. ("no” blinks)
4. Press the PRINT key to switch to
ClrC go
. ("go” blinks)
□
To cancel the operation, press the ZERO key.
You will be returned to the status in step 2.
5. If
ClrC go
is displayed, press the TARE key to
restore to factory default and display
end
.
To end setting, press the ZERO key or CAL switch. This
will return you to weighing mode.
Press
(2 times)
Press
Press