40
□ To set the HI limit value (no response if
erCd 0
)
Sends a six-digit number that does not contain “+/-” or decimal point
Response H I : + 0 0 0 5 0 0 C
R
L
F
□ To set the LO limit value (no response if
erCd 0
)
Sends a six-digit number that does not contain “+/-” or decimal point
Command L O : + 0 0 0 2 0 0 C
R
L
F
Response L O : + 0 0 0 2 0 0 C
R
L
F
□ To set the comparator memory (no response if
erCd 0
)
Command C N : X X C
R
L
F
XX is memory number (00
~
09
)
Response C N : X X C
R
L
F
□ To output the HI limit value
Command ? H I C
R
L
F
Response H I , + 0 0 0 5 0 0 C
R
L
F
□ To output the LO limit value
Command ? L O C
R
L
F
Response L O , + 0 0 0 2 0 0 C
R
L
F
□ To recall the comparator memory
Command ? C N
Response C N , X X C
R
L
F
XX is memory number (00
~
09
)
□ The scale is not in a state where a command could be executed.
Then, the scale will reply “I”.
Command Z C
R
L
F
Response I C
R
L
F
The scale is not in a condition that zero operation is possible.
□ Command does not exist for the scale. Then, the scale will reply “?”.
Command B C
R
L
F
Response ? C
R
L
F
The scale received an undefined command.
Command H I : + 0 0 0 5 0 0 C
R
L
F