10
6.2. Operation Keys
Operation key
Description
ON/OFF Key
Shows or hides the display.
Note: When the power plug of the scale is connected, power is supplied
to the electric circuits inside the scale.
SAMPLE Key
Selecting “pcs” as the unit activates individual weight setting mode.
Pressing and holding activates comparator threshold setting mode.
PRINT Key
Outputs the weighed value as data.
Increases the value of the digit flashing when setting.
MODE Key
Switches the weighing unit.
Shifts to the right the digit flashing when making settings.
ZERO Key
Performs zeroing operation. (See below.)
TARE Key
Performs tare operation. (See below.)
Saves the set value and proceeds to the next item when setting.
Zeroing operation
With the zero point gained when turning on the display (power on zero function) as reference,
pressing the ZERO key when the weighed value is stable within ±2% of the weighing
capacity sets the zero point there and the zero point indicator ◄ is lit
when zero is displayed.
If tare function is in use, the net is deleted.
Tare function
Pressing the TARE key when the weighed value is a stable positive value subtracts the
weight of the item on the weighing pan as the weight of the container (tare) weight. The
displayed value becomes zero value and both the zero point indicator ◄ and net indicator ◄
are lit.
If the container (tare) is removed from the weighing pan while the tare function is in use and
the value returns to the zero point, net is displayed as a negative value.
Note: The weighing range is reduced by the amount of the tare mass.
Power on Zero Function
Power on zero is a function that displays zero along with the zero point indicator
◄
when the
display is turned on by the ON/OFF key with nothing on the weighing pan.
The power on zero function can be used within a range ±50% of weighing capacity from zero
during sensitivity adjustment.
Note: If a tare or similar device is used on the weighing pan before weighing starts (power on),
weighing performance may be affected.