39
Error code
Description and how to resolve
EC,E00
Communications error
A protocol error occurred in communications. Check the format and the
baud rate.
EC,E01
Undefined command error
An undefined command was received. Check the command.
EC,E02
Not ready
The command received cannot be processed.
e.g. The scale received a Q command, which requested the weighing
data, but it was not in weighing mode with the display on.
e.g. The scale received a Q command while processing a RE-ZERO
command.
Adjust the timing of transmitting the command.
EC,E06
Format error
The format of the received command is incorrect.
e.g. The data is numerically incorrect.
e.g. Alphabet characters are input instead of values.
Check the command.
EC,E07
Setting value error
The received data exceeds the range of values that the scale can accept.
Check the parameter values range of the command.
The following is an example for function table
erCd 2
(Echo Back Response).
□
To zero the weighed value when it is stable (no response if
erCd 0
)
Command Z C
R
L
F
Response Z C
R
L
F
When zeroing operation is possible
□
To tare the weighed value when it is stable (no response if
erCd 0
)
Command T C
R
L
F
Response T C
R
L
F
When zeroing operation is possible
□
To switch the weighing unit (no response if
erCd 0
)
Command U C
R
L
F
Response U C
R
L
F
When zeroing operation is possible
□
To set the ID number (no response if
erCd 0
)
Command I D : 0 0 0 1 0 0 C
R
L
F
Response I D : 0 0 0 1 0 0 C
R
L
F
□
To request the ID number
Command ? I D C
R
L
F
Response I D , 0 0 0 1 0 0 C
R
L
F