Calibration Test Report
Calibration test mode is used to compare a calibration weight with the calibration test data
weighed by the scale.
□
This test does not perform sensitivity adjustment.
1. In weighing mode, press and hold the CAL switch until
1
C
CC
appears, and then release the switch, or, press and
hold the TARE key for five seconds until
CC
appears
and release the key.
□
Calibration test mode is not available when the function
setting
info 0
is selected.
2. Press the TARE key to display
CC 0
.
3. Change the weight value if necessary. To change the weight
value, press the MODE key. Change the value using the
following keys.
4. With nothing on the weighing pan, press the TARE key.
The zero point is measured and the weighed value is
displayed in “kg” for a few seconds. Then, the value of the
weight is displayed.
5. Place a weight of the displayed value on the
weighing pan and press the TARE key to
measure it. The weight value is displayed in kg for
a few seconds.
6.
end
appears.
7.
gmp
appears and the calibration test report is output.
8.
CC
appears again. Remove the weight.
Press the ON/OFF key to turn the display OFF or press
the CAL switch.
Keys
Content
Moves to the next item.
Increases the blinking digit by 1.
Sets the parameter.
Weight
CAL switch
Release the key
Press
Change the weight
value if necessary
Weighing Pan
Weighing unit
Press
Press
Weighing unit
To output the results