EN
29
Conducted Radio Frequency
Disturbance Test – CS.
IEC/EN 61000-4-6
3Vrms
150kHz to 80MHz
0.15 – 80MHz, 3Vrms, 80%
AM, 2Hz
Portable and mobile RF communications equipment
shall be used no closer to any part of the aScope 3
system, including its cables, than the recommended
separation distance calculated from the equipment
of the frequency of the transmitter.
Recommended Seperation Distance D = 1.17√P
Radio Frequency Electromagnetic
Field Susceptibility Test – RS
IEC/EN 61000-4-3
3V/m; 80Hz to 2.5GHz
80-2500 MHz, 3V/m, 80%
AM(2Hz)
Note 1: The guidance may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
a) Field strengths from fixed transmitters, such as base stations for radio (celluar/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast, and TV broadcast, cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmit-
ters, an electromagnetic site survey shall be considered. If the measured field strength in the location in which the aScope 3 system is used exceeds the
applicable RF compliance level above, the aScope 3 system shall be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the aScope 3 system unit.
b) Over the frequency range 150kHz to 80MHz, field strengths shall be less than 3V/m
Recommended Separation Distances Between Portable and Mobile RF Communication Equipment and aScope 3 system.
The aScope 3 system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The user of the aScope 3
system can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters and the aScope 3 system as recommended below, according to the maximum output power of the communication equipment.
Rated maximum output power (W)
of transmitter
Separation distance (m) according to frequency of transmitter
150kHZ to 80MHz
D = 1.17√P
80MHz to 800MHZ
D = 1.17√P
800MHz to 2.5GHz
D = 2.33√P
0.01
0.12m
0.12m
0.23m
0.1
0.37m
0.37m
0.74m
1
1.17m
1.17m
2.33m
10
3.70m
3.70m
7.37m
100
11.70m
11.70m
23.30m
For transmitters rated at a maximum output power not listed above, the recommended separation distance (D) in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note 1: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.