background image

EN

12

Preparation of aScope 3

1.  Carefully slide the bending control lever forwards and backwards in each direction until it stops. Confirm that the bending section functions smoothly 

and correctly.

2. Slide the bending lever slowly to its neutral position. Confirm that the bending section returns smoothly to a neutral position.
3.  Using a syringe insert 2ml of sterile water into the working channel port (if Luer Lock syringe use the enclosed introducer). Depress the plunger, ensure 

there are no leaks, and that water is emitted from the distal end.

4.  If it is anticipated that suction will be required during the procedure, prepare the suction equipment according to the supplier’s manual. Connect the 

suctioning tube to the suction connector and press the suction button to check that suction is applied.

5.  If applicable, verify that endoscopic accessory of appropriate size can be passed through the working channel without resistance. The enclosed  

introducer can be used to facilitate the insertion of soft accessories such as microbiology brushes.

3.2. Operating the aScope 3

WARNING  

•  Excessive force should never be used when operating aScope 3.
•  If any malfunction should occur during the endoscopic procedure, stop the procedure immediately, put the distal tip in its neutral and non-angled  

position and slowly withdraw the aScope 3 without touching the bending lever. 

•  Always observe the live endoscopic image while withdrawing the aScope 3.  
•  If needed remove secretion or blood from the airway before and during the procedure. The suction function of any appropriate suction device can be used 

for this purpose. 

•  The temperature of the distal end of the endoscope may reach up to 45C (113F) due to heating of the LEDs. Long, sustained contact with the mucosal  

membrane may cause mucosal injury. Avoid long periods of contact between the tip of the device and the mucosal membrane. 

Holding the aScope 3

The handle of the aScope 3 can be held in either hand. Use the thumb to move the control lever and the index finger to operate  
the suction button. The hand that is not holding the aScope 3 can be used to advance the insertion cord into the patient’s mouth  
or nose.

Manipulating the tip of the aScope 3

The control lever is used to flex and extend the tip of the aScope 3 in the vertical plan. Moving the control lever downward will make 
the tip bend anteriorly (flexion). Moving it upward will make the tip bend posteriorly (extension). 

The insertion cord should be held as straight as possible at all times in order to secure an optimal tip bending angle.

CAUTION

Do not exert excessive force on the bending section as this may result in damage to the equipment. Examples of inappropriate 
handling of the bending section include:
- Manual twisting
- Operating it inside an ETT or in any other case where resistance is felt
- Inserting it into a preshaped tube or a tracheostomy tube with the bending direction not aligned with the curve of the tube

Tube connection

The tube connection can be used to mount ETT and DLT with an ISO connector during intubation.

Summary of Contents for aScope 3 5.0/2.2

Page 1: ...Instructions for use Ambu aScope 3 5 0 2 2 aScope 3 Slim 3 8 1 2 Ambu aView For use by trained clinicians physicians only For in hospital use...

Page 2: ...Fax 44 0 1480 498 405 www ambu co uk Italy Ambu S R L Via Paracelso 18 Centro Direzionale Colleoni 20864 Agrate Brianza MB Italia Tel 39 039 657811 Fax 39 039 6898177 www ambu it Spain Firma Ambu S L...

Page 3: ...ati tmutat 308 335 Italiano Manuale d uso 336 363 364 391 Lietuvi kai Naudojimo instrukcijos 392 418 Latviski Lieto anas instrukcija 419 446 Nederlands Gebruiksaanwijzing 447 474 Norsk Brukerveilednin...

Page 4: ...stem Usage 11 4 Cleaning and Disinfection of aView 15 5 Technical Product Specifications 15 6 Connecting accessories 19 7 How to Operate aView 21 8 Trouble Shooting 32 Appendix 1 Electromagnetic Compa...

Page 5: ...is a single use device designed for use in adults and children It has been clinically evaluated for the following minimum endotracheal tubes ETT and double lumen tubes DLT sizes Minimum ETT inner dia...

Page 6: ...Be careful not to damage the insertion cord or distal tip when using sharp devices such as needles in combination with the aScope 3 Be careful when handling the distal tip of the insertion cord and do...

Page 7: ...connection Unless specified otherwise the text relates to both products In this document the term aScope 3 always refers to both variants If one of the variants is referred to specifically the terms...

Page 8: ...on the bracket Bracket Secures the monitor to e g an IV pole Pouch hook Slide the hook through the 3 holes on the bracket It can now be used to hold the aScope 3 pouch hole in top corner for easy stor...

Page 9: ...tion between 80 and 109 kPa in operating environment Connection to external monitor Direct current Alternating current Symbol of Class II equipment IP30 Protection against solid objects Waste Bin symb...

Page 10: ...EN 10 Lot Number Batch Code Serial Number Warning Consult Instruction for use Year of Manufacture followed by YYYY...

Page 11: ...a charger be readily available during use Place or hang the aView display on a stable support while in use Dropping aView could damage it Position the power cord where it is unlikely to be stepped on...

Page 12: ...awing the aScope 3 If needed remove secretion or blood from the airway before and during the procedure The suction function of any appropriate suction device can be used for this purpose The temperatu...

Page 13: ...nnector is connected to a suction device Apply a vacuum of 85 kPa 638 mmHg or less when suctioning Applying too large a vacuum may make it difficult to terminate suctioning CAUTION Secure the tubing p...

Page 14: ...r use and must be disposed of in accordance with local guidelines for collection of infected medical devices with electronic components Do not soak rinse or sterilize this device as these procedures m...

Page 15: ...View must be submitted to the pre check procedure in section 5 1 Between procedures aView must be stored in accordance with local guidelines 5 Technical Product Specifications 5 1 aScope 3 Specificati...

Page 16: ...D Brightness control Yes Contrast control Yes Start up time About 1 second Memory Storage capacity 8GB Electrical power Power requirement 18V 1 67A DC input Battery type 11 1V 3760mAh Operating enviro...

Page 17: ...shall be discarded aView can be placed on a solid flat surface by using the stand on the back of aView If needed aView can be placed by using the supplied bracket 6 1 Charging aView Connect the aView...

Page 18: ...LE MODE SIMPLE MODE Live image available and the User Interface displays basic user functions Advanced functions can be initiated from SIMPLE MODE by pressing ADVANCED MODE button aView loads the User...

Page 19: ...recorded image NOTE When de selecting a live image by leaving the blue Live Image tab to go to an other tab in ADVANCED MODE aView asks for confirmation Live image vs recorded image in ADVANCED MODE W...

Page 20: ...or a blue screen will appear After about one minute aView will be ready in Simple Mode NOTE During startup the screen may flicker for about a second How to turn off aView Assumption aView is powered...

Page 21: ...lour of the device interface where the Ambu Visualization Device is connected How to take a SNAPSHOT Accessible in ADVANCED MODE blue Live Image tab Keep the Ambu Visualization device as still as poss...

Page 22: ...tion Snapshot has already been located Press the snapshot to select Press to view the snapshot Press or to toggle between snapshots in the folder NOTE Double click to Select and View How to view a vid...

Page 23: ...les are not deleted from aView in the copy process How to set the time and date Accessible in ADVANCED MODE Green Settings tab Press to access system time and date Press to toggle between 12 and 24 ho...

Page 24: ...e ADMIN account is logged in Connect the power adapter before commencing a software update Press to enter software settings Insert USB stick containing the upgrade file in its root directory Press to...

Page 25: ...ED MODE Green Settings tab Press to enter Display settings Press the up or down arrow to adjust the colour for the Red Green and blue The arrow will light up red when pressed and the achieved level wi...

Page 26: ...A communication error has occurred between aScope 3 and aView Restart the system by pressing the ON OFF button on aView for at least 2 seconds When aView is OFF restart by pressing ON OFF button once...

Page 27: ...channel Remove endoscopic accessory Difficult to insert endoscopic accessory through the channel Cause Action Channel blocked Clean the working channel using a cleaning brush or flush the working cha...

Page 28: ...issions IEC EN 61000 3 2 N A N A Voltage Fluctuations Flicker emissions IEC EN 61000 3 3 Complies Guidance and Manufacturer s Declaration Electromagnetic Immunity The aScope 3 system is intended for u...

Page 29: ...tional measures may be necessary such as reorienting or relocating the aScope 3 system unit b Over the frequency range 150kHz to 80MHz field strengths shall be less than 3V m Recommended Separation Di...

Page 30: ...lateral standard Electromagnetic compatibility Requirements for test The Ambu aView power supply conforms with Council Directive 93 42 EEC concerning Medical Devices IEC 60601 1 Medical electrical equ...

Reviews: