Under certain conditions, E
ASY
H
EAT
may suggest changing your capacitor. Call Ambrell
service to order a replacement capacitor, then follow these instructions:
Do this...
Observe...
turn off AC power switch at rear
AC power LED, panel are not lighted
remove line cord from rear of unit
use Philips screwdriver to remove
screws securing cover; remove cover
note location of capacitor
with a 5mm hex wrench, remove 2
screws securing capacitor; remove
apply thin coat of thermal grease to
mounting surface of new cap
use screws to install new cap (no
polarity); tighten
replace cover, secure with screws;
replace cord and check cooling water
turn on AC power switch
Navigate to Control zone:
►
RF-Set
select
►
Cap-set
and select the new
value (call Ambrell for help)
returns to Home zone
Tap Cap Val(uF)
►10 1x0.66 =0.66