Introduction
11
801-9309.doc
Insert drain tube end fully into fixture labeled OUTLET
(To remove tubing, press retaining ring while pulling on tube.)
Flow monitoring
The amount of water flowing through your equipment is monitored, since inadequate or
absent cooling water can result in damage to the coil or capacitor. When flow is adequate
to the heat sink and to the coil – 0.7 g/m (2.6 l/m) or more total – two internal switches
are closed. When flow falls off to the heat sink or the coil, one or both of the switches
open and the power supply senses it, halts RF power and indicates
Ω
Fault.
No action required by you; internal connections.
Line cord
Your EASYHEAT
power supply accepts line-voltage through this cord. The power
switch module includes a circuit breaker rated for your equipment.
assure unit cover is attached and secure; assure power switch is OFF
insert line cord at rear of unit
insert line cord into grounded outlet
turn power switch ON
Start-up display
Upon power-up, certain important information about your EASYHEAT is displayed:
Output power capability: Depending on the unit
you purchased,
1kW RF Power
or
2kW RF
Power
Firmware versions: There may be instances where
we need this information from you.
n.nn
represents the version numbers of the Display or
Main firmware
SYS ID #: The SYS ID# is used to connect to
remote monitoring software. Refer to § 3.4 for
more information on eView Induction Heating
Software.
You can also view the output power capability screen and the Firmware version screen
when the unit is ON; while in the
Control Zone
, §2.2, press
▲
or
▼
to repeat the
Start-up display screens
The next section explains what you see when the unit is operating in the
Home
zone.
EASYHeat 1012
1kW RF Power
Display: Vn.nn
Main: Vn.nn
Certain cooling configurations can significantly restrict flow, limiting
EASYHEAT’s efficiency. Under these conditions, you must supply water to the unit
from 3/8” tube and adapt to ¼” at the unit if:
•
supply line lengths exceed 5 feet
•
EASYHEAT is in series with another power supply or component
SYS ID # = nnnnn