Problem
Possible Cause /
Corrective Action
Audible alarm cannot be
temporarily silenced using
the Test/Reset button.
The temporary disable timer is set to
0. See Configuration Parameter 03 to
set the temporary disable timer.
Monitor is flashing “diS.”
The monitor is in night set back mode
and is disabled. See Configuration
Parameter P07.
Monitor keeps bouncing
back and forth between
adjacent zones.
The monitor is being influenced by an
external air source. Remove the
source of drafts. See Configuration
Parameters P04 and P05 to set the
transition delays.