30
30
If the monitor is configured for open contacts to activate the external
alarm, the display will show:
If the monitor is configured for closed contacts to activate the sash
override alarm, the display will show:
If the monitor is configured for open contacts to activate the sash
override alarm, the display will show:
Press either the Up or Down button to scroll through the settings.
When the desired setting is displayed, press the Test/Reset button. The
PGM descriptor will flash once and the horn will give two quick beeps
to acknowledge the save.
The monitor will return to the P08 Configuration menu selection. Press
the Up and Down Buttons to advance to another Configuration
Parameter. Press and hold the hold the Test/Reset button for 2 seconds
to exit the Configuration menu.
P09 — Relay Output 1
This relay output has contacts that are open in the non-energized state.
The contacts can be configured to close when the function is activated
or to open when the function is activated or power is lost. It may also
be deactivated. The table below describes the selections: