28
28
low alarm setting, but does not allow you to turn it off. See P12 for
information on turning the night setback low alarm off.
When the desired setting (Remote Test/Reset, Remote Emergency
Purge, or Night Setback) is displayed, press the Test/Reset button. The
PGM descriptor will flash once and the horn will emit two quick beeps
to acknowledge that the setting has been saved.
The monitor will return to the P07 Configuration menu selection. Press
the Up and Down Buttons to advance to another Configuration
Parameter. Press and hold the hold the Test/Reset button for 2 seconds
to exit the Configuration menu.
P08 — Alarm Input
This Configuration Parameter configures the switch settings for the
instrument’s Remote Test/Reset , Remote Emergency Purge, and Sash
Override/Alarm Input features. After this Configuration menu selection
is entered, the monitor will turn on the Program mode PGM descriptor
and display the current setting. Press the Up and Down arrow keys to
scroll through the available settings.
NOTE:
P08 may only be configured to have one of these features
operational at a time. The Remote Test/Reset and Remote Emergency
Purge functions are also available under P07.
Remote Test/Reset
— This allows the operator to activate the
instrument’s Test/Reset function remotely via a contact closure. It
functions in the same manner as the Test/Reset button on the
instrument’s front panel (see Section 1). The local Test/Reset button is
still active even if the instrument is configured for Remote Test/Reset.
If this function is active, the following display will appear under P08:
Remote Emergency Purge
— This allows the operator to active the
instrument’s Emergency Purge function remotely via a contact closure.
It functions in the same manner as the Emergency Purge button on the
instrument’s front panel (see Section 1). The local Emergency Purge
button is still active even if the instrument is configured for Remote
Emergency Purge. If this function is active, the following display will
appear under P08: