20
20
Configuration menu after 1 minute has elapsed without any keypad
activity.
SUMMARY:
Pressing the Test/Reset button for 10 seconds accesses
the Configuration menu. Pressing the Up or Down Button advances to
the next parameter and also navigates within a selection. Pressing
once manually scrolls; holding scrolls quickly. Pressing and releasing
the Test/Reset button after a change has been made saves it; pressing
and holding the Test/Reset Button for two seconds cancels the change
and exits.
Configuration Parameters
The following table shows the factory default settings for the monitor’s
various programmable parameters. A default reset restores these
settings.
Configuration Parameter
Factory Default
CAL — Calibration
(A field calibration is required.
See the Section 3: Calibration)
P01 — Digits enabled/disabled
Disabled
P02 — Units of measure
English (feet/minute)
P03 — Temporary horn disable
timer
255 (infinite)
P04 — Caution-to-alarm
transition delay timer
3 seconds
P05 — Alarm-to-caution
transition delay timer
3 seconds
P06 — Low alarm caution offset
20 fpm
P07 — Night setback input
CL.1 (night setback mode active
with contact closure)
P08 — Alarm input
CL.2 (sash over-ride, activates
with contact closure)
P09 — Relay output 1
CL.1 (alarm output, contacts
close with alarm)
P10 — Relay output 2
OP.2 (alarm output, contacts
open in alarm)
P11 — High alarm relay
OFF
P12 — Night setback low alarm
OFF
dEF — Default reset
Resets P01 – P12 parameters to
factory defaults