80
Allen® Advance Table User Manual (D-720686-A6)
Electromagnetic Immunity Guidance
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The A-71100-XX/A-71101-XX and A-71705-XX are intended for use in the electromagnetic
environment specified below. The customer or the user of the A-71100-XX/A-71101-XX or
A-71705-XX should make sure it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic
Environment—
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 6 kV Contact
± 8 kV Air
± 6 kV Contact
± 8 kV Air
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
± 2 kV for Power
Supply Lines
± 1 kV for Input/
Output Lines
± 2 kV for Power
Supply Lines
± 1 kV for Input/
Output Lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV Line(s) to
Line(s)
± 2 kV Line(s) to
Earth
± 1 kV Line(s) to
Line(s)
± 2 kV Line(s) to
Earth
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage Dips, Short
Interruptions, and
Voltage Variations
On Power Supply
Input Lines
IEC 61000-4-11
< 5% U
T
(> 95% dip in U
T
)
for 0.5 cycles
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in U
T
) for
25 cycles
< 5% U
T
(> 95% dip in U
T
)
for 5 s
< 5% U
T
(> 95% dip in U
T
)
for 0.5 cycles
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in U
T
) for
25 cycles
< 5% U
T
(> 95% dip in U
T
)
for 5 seconds
Mains power quality
should be of a typical
commercial or hospital
environment. If the user
of the A-71100-XX/A-
71101-XX or A-71705-XX
requires continued
operation during power
mains interruption, it is
recommended that the
A-71100-XX/A-71101-XX
or A-71705-XX be
powered from an
uninterruptedly power
supply or a battery.
Power Frequency
(50/60Hz)
Magnetic Field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields should
be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Note: U
T
is the AC mains voltage prior to application of the test level.