42
Allen® Advance Table User Manual (D-720686-A6)
NOTE:
We recommend that you use the two large security straps to secure the
patient to the support top.
•
The Table is free of objects that could interfere with the adjustment.
2.
If the Table is in a Trendelenburg position, press and hold the
opposite
Trendelenburg
control until the lowered column is at the
same height as the raised column and the
Tilt Level
indicator on the
display is
green
.
NOTE:
Near the bottom on the outside of the columns,
there are numbers to help identify the height
level.
3.
If the support top is tilted, press and hold the
opposite
Tilt
control until the top is no
longer tilted. The
Tilt
indicator on the display
should be in the center of the Tilt Indicator Area, and the
Tilt Level
indicator should be
green
.
R
OTATE
THE
S
UPPORT
T
OP
90°
WARNING:
Before you adjust the Table or use the flip or tilt features, make sure that
the brakes are locked, the patient is secured against the prone support
set, the patient is fully prepared for the adjustment; the Table is free of
objects that could interfere with the adjustment, and all loose or
unsupported objects have been secured. Otherwise, patient fall, injury, or
equipment damage could occur.
1.
Make sure that—