![allen A-71100-BR User Manual Download Page 49](http://html.mh-extra.com/html/allen/a-71100-br/a-71100-br_user-manual_2898489049.webp)
Allen® Advance Table User Manual (D-720686-A6)
41
NOTE:
We recommend that you use the two large security straps to secure the
patient to the support top.
•
The Table is free of objects that could interfere with the adjustment.
WARNING:
Observe lines closely during articulations. Always use good line
management techniques. Failure to do so could cause patient fall, injury,
or equipment damage.
2.
Press and hold the applicable
Tilt
control until
the support top is at the correct angle. When
you release the control, the support top will
lock at that angle.
The Tilt indicator on the display will
show how far the support top is tilted to
the left or right.
3.
To return the support top to the flat
position, press and hold the opposite
Tilt
control until the top is no
longer tilted. The Tilt indicator on the display will be in the center of
the Tilt Indicator Area and the Tilt Level indicator will be green.
L
EVEL
THE
T
ABLE
AND
THE
S
UPPORT
T
OP
To determine if the Table is level, look at the display:
•
The
Tilt Level
indicator will be green.
•
The
Tilt
indicator will be in the center of
the Tilt Indicator Area.
WARNING:
Before you adjust the Table or use the tilt feature, make sure that the
brakes are locked, the patient is secured to the support top and is fully
prepared for the movement, and the Table is free of objects that could
interfere with the adjustment. Otherwise, patient fall, injury, or equipment
damage could occur.
1.
Before you adjust the Table, make sure that—
•
The brakes are locked (the
Locked
indicator on the
Table display is
green
).
•
The patient is secured to the support top and is fully
prepared for the adjustment.