58
Allen® Advance Table User Manual (D-720686-A6)
5.
Pull the assembly out from the
column, and make sure the
assembly is in the locked-out
position. The Cervical Traction
Release button will return to its
original position.
WARNING:
The maximum diameter of the
traction rope should be 0.25" (6.35
mm). The use of a traction rope that
is greater than 0.25" (6.35 mm) in
diameter could cause binding inside the cervical traction assembly.
Patient injury could occur as a result of inaccurate traction.
6.
Put the traction rope through
the carriage, and through the
hole in the center of the
bowtie. The rope will go
through the traction assembly
and out the left side of the
column, near the release.
7.
Put the traction rope over and
around the pulley.
8.
Press the carriage adjustment
releases and put the carriage in
the desired position. See
“Change the Cervical Traction
Angle” on page 59.
WARNING:
The maximum weight load for the
traction assembly is 50 lb (22.6 kg).
To exceed this weight could cause