Test
Standard
Class / Severity level
Test results
Radiated emissions
in freq. range 30 - 6000 MHz
ETSI EN 301 489-1;
ETSI EN 301 489-17 /
EN 55032
Class B
Complies
Harmonic current test
ETSI EN 301 489-1;
ETSI EN 301 489-17 /
EN 61000-3-2
230 VAC mains
Complies
Flicker tests
ETSI EN 301 489-1 /
EN 61000-3-3
230 VAC mains
Complies
Immunity
(per ETSI EN 301 489-1, ETSI EN 301 489-17)
Immunity from Electrostatic
discharge (ESD)
EN 61000-4-2
4 kV contact discharge
8 kV air discharge
Complies
Immunity from radiated
electromagnetic fields
EN 61000-4-3
3.0 V/m,
80 MHz ¸ 6.0 GHz,
80% AM, 1 kHz
Complies
Immunity from Electrical
Fast transient (EFT)
EN 61000-4-4
AC mains: ± 1.0 kV;
Tr/Th – 5/50 ns, 5 kHz
Complies
Immunity from Surge
EN 61000-4-5
AC mains: ± 1.0 kV DM / ± 2.0 kV CM,
Tr/Th – 1.2/50 (8/20) ms
Complies
Immunity from conducted
disturbances induced by
radio-frequency fields
EN 61000-4-6
AC mains: 3.0 VRMS;
0.15÷80 MHz, 80% AM @ 1 kHz
Complies
Immunity from Voltage
interruptions
EN 61000-4-11
AC mains:
0% - 0.5 cycle & 1 cycle;
70% - 25 cycles; 0% - 250 cycles
Complies