11.3
Cleaning and disinfecting the cradle
The wand cradle requires the procedures in the following sections for cleaning and disinfection.
These procedures must be carried out:
l
After the scanner assembly, before first-time use
l
Between patients
Warning:
Avoid deviating from the recommended cleaning and disinfection guidelines, and modifying or
substituting recommended materials, to prevent biological hazard.
You must follow all the cleaning and disinfecting steps below to ensure that the cradle is properly reprocessed and
ready for use.
11.3.1 Preparation before cleaning and disinfection
1. Visually inspect the cradle for any noticeable damage, for example, deterioration such as discoloration, pitting,
or cracks.
CAUTION:
Do not clean, disinfect, or use the cradle if any damage is detected. Please contact iTero Customer
Support for further instructions.
2. Prepare the following:
o
Required cleaning and disinfecting materials:
l
CaviWipes1 (or, for a list of alternative materials and the required contact time, see
)
l
Soft bristle brush (e.g. the smaller end of a Healthmark Trumpet Valve Brush 1mm diameter,
Cat # 3770 or equivalent)
o
PPE and work environment
l
Please follow the cleaning and disinfecting material manufacturers’ instructions.
Note:
Replace cleaning and disinfection materials (brushes/wipes) if visibly damaged or soiled.
Before starting the cleaning and disinfection procedure, put on your PPE.
iTero Element 5D and iTero Element 5D Plus imaging systems User manual
© 2022 Align Technology, Inc. All rights reserved.
187
Summary of Contents for iTero Element 5D
Page 1: ...it starts with iTero Element 5D and iTero Element 5D Plus imaging systems User manual...
Page 2: ......
Page 245: ......