MAINTENANCE
51
MAINTENANCE
Accuracy of fluid delivery is determined by measuring the
volume of fluid delivered over a known time period and
comparing this to the expected value. To ensure accurate
measurements during the test, a volumetric glass burette (class
A) must be used to collect the fluid. The infusion time interval
must be 180 seconds or greater to minimize measurement
errors. During a 180-second test, a one-second error by the
operator results in an error of 0.6%.
A.
Test Equipment Setup
1.
Obtain a new administration set and connect it to a 500-
milliliter container which is at least half full. Prime the set
and eliminate all air.
2.
Connect the apparatus as shown in Test Equipment Setup
below. Use a volumetric burette marked in 0.1-milliliter
increments (class A glassware).
3.
Install the cassette to the channel to begin test.
New Instrument Volume Accuracy Test
HEAD HEIGHT
FROM DRIP
CHAMBER
SOLUTION IN 500-ml CONTAINER
(HALF FULL AT BEGINNING OF TEST)
60cc SYRINGE
STOPCOCK
FLUID LEVEL AT END OF TEST
(READ AT BOTTOM OF
MENISCUS)
FLUID LEVEL AT START OF
TEST
(READ AT BOTTOM OF
MENISCUS)
Test Equipment Setup
Watchdog Audio Test
Manually test the Watchdog Alarm Audio from the Standard
Display. Press the
More Options
button two times until a softkey
labeled Demo WD appears. Press this key and follow the
directions on the screen for completing the Watchdog Test.
Summary of Contents for MedSystem III
Page 37: ...34 GETTING STARTED T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K...
Page 45: ...T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K 42 ALARMS ADVISORIES AND PROMPTS...
Page 61: ...58 MAINTENANCE T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K...
Page 63: ...T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K 60 MAINTENANCE...