INTRODUCTION
Features
Multi-channel Fluid Delivery System
The instrument combines three independent infusion
channels in an unparalleled small size.
Lightweight/portable
The pump with pole clamp weighs just over 5 pounds and
is easy to transport.
Unique, rotating pole clamp
The pump may be attached to a variety of surfaces.
Dose Rate Calculator (DRC)
The pump calculates a volumetric or dose rate based on
values entered for patient weight, drug concentration (drug
amount and diluent volume) and dosing parameters.
Drug List Editor (DLE)
The pump drug list can be customized using Drug List
Editor software.
Six Device Types available
Six available Device Types with configurable parameters
(maximum and minimum rates, maximum volumes,
baseline and maximum pressures, and air-in-line thresholds)
to achieve specific clinical applications:
Free-flow Protection
The MedSystem III® system Administration Sets contain a
cassette that provides protection from free-flow conditions.
To remove the cassette from the pump, the cassette’s slide
clamp is pulled to full extension, occluding the tubing and
preventing fluid from flowing.
General Purpose
Neonatal
Controller Pressure
Operating Room
General Purpose II
Operating Room II
B
C
A
START
STOP
MORE
OPTIONS
STANDARD
DISPLAY
O N / O F F
®
®
DLE
®
A: Stopped
DRUG?
Wt ---- KG= 0.0 LB
A:Conc ------- mg/---- ml
A:Rate ------- ml/h
A:Dose ------- mcg/kg/min
Press Select to choose line
Select
Clear
A:VR 1 ml (Vol Rem)
A:VI 0 ml DI 0.0 mg
Summary of Contents for MedSystem III
Page 37: ...34 GETTING STARTED T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K...
Page 45: ...T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K 42 ALARMS ADVISORIES AND PROMPTS...
Page 61: ...58 MAINTENANCE T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K...
Page 63: ...T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K 60 MAINTENANCE...