GET
TING ST
ARTED
GETTING STARTED
23
5.
Press to begin infusion.
• Four tones sound (if primary infusion is in progress).
• Pump starts infusing at secondary rate.
• Current date and time are entered.
6.
Press
OR
• Display reverts to Standard Display page after one
minute.
7.
Verify settings.
8.
Verify solution flow from secondary container.
The infusion rate can be set with the volume remaining (VR) and
time entry.
1.
Press from the Standard Display.
2.
Press
C
Co
on
nf
fi
ig
g
softkey at bottom of display.
3.
Select C
Ch
ha
an
ng
ge
e IIn
nssttrru
um
me
en
ntt S
Se
ettttiin
ng
gss from menu.
4.
Press
↑↑
or
↓↓
, softkey to change Setup Line Option from N
NO
O
to Y
YE
ES
S..
5.
Return to the Standard Display and press a channel.
• The display will read “Setup: Select VR and Rate”.
6.
Step down to “Set: Select VR and Rate”.
7.
Press
A
Ac
cc
ce
ep
pt
t
to set a time.
1.
Press ,
or
.
• Secondary programming page is displayed.
• Rate is highlighted.
2.
Press
↑↑
,
↓↓
,
F
Fa
as
st
t
↑↑
or
F
Fa
as
st
t
↓↓
to change rate.
• Value flashes.
3.
Press
E
En
nt
te
er
r
to confirm.
• New rate begins infusing immediately.
1.
Press ,
or
.
• Secondary programming page is displayed.
2.
Press .
•
1
1
°°
P
Pr
ri
i
and
C
Ca
al
lc
cO
On
n
softkeys appear.
3.
Press
1
1
°°
P
Pr
ri
i
softkey.
• Primary programming page is displayed.
Secondary Mode (Continued)
To set secondary rate (Continued)
To titrate or change secondary rate during infusion
To review or change primary value(s) during secondary infusion
Secondary
100
Infusing
Stop affects channel A
Cntrst GP
VR = 91
VI = 9
ml/h
NOTE:
Channel display on the Standard Display is reverse
highlighted.
A: Secondary Rate 100 ml/h
A: Infusing Secondary
A: Sec VolRem (VR)
50
ml
A: Pri VolInf(VI) 200 ml
A: Sec Time(TR) -----
since 12:37p 01 Feb 02
Stop affects Secondary
Select
Repeat
To set rate of infusion from a time entry
A: Pri VolInf(VI) 1 ml
A: Primary Rate 100 ml/h
A: Stopped
A: Pri VolRem (VR) 500 ml
A: Pri Time(TR) 05h 00m
since 02:16p 01 Jul 02
Press Select to choose line
Select
Fast
Summary of Contents for MedSystem III
Page 37: ...34 GETTING STARTED T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K...
Page 45: ...T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K 42 ALARMS ADVISORIES AND PROMPTS...
Page 61: ...58 MAINTENANCE T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K...
Page 63: ...T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K 60 MAINTENANCE...