ALARMS, ADVISORIES AND PROMPTS
39
Alarms (Continued)
ALARMS, ADVISORIES
AND PROMPTS
Alarm
Meaning
Response
Cassette Not Latched
Push cassette completely in.
Ensure slide clamp is flush with
entire cassette. Press to
resume infusion.
If condition continues, try
cassette in a different channel.
Replace administration set if
alarm recurs.
Clean lower assembly according
to cleaning instruction described
in MAINTENANCE section of this
document.
Cassette Removed
Reinstall cassette, and press
to resume infusion
OR
Press Cancel.
Check Fluid Side
Check tubing between container
and pump for a closed regulating
clamp, closed vent (with
unvented container), kinked
tubing, empty syringe, or any
restriction to flow.
If NO occlusion is present, press
CONFIRM.
Press to resume infusion.
Verify fluid is flowing in drip
chamber.
A
4
appears on standard
display to indicate Confirm has
been pressed.
Cassette is partially disengaged or
latching mechanism is dirty.
Cassette is removed from holder
while channel is infusing.
Pos sible upstre am restrictions to
flow.
Faulty Cassette
Reinsert cassette in another
channel.
If alarm recurs in second channel,
replace administration set.
If alarm recurs with two cassettes
in the same channel, discontinue
use and contact qualified service
personnel.
Cassette may be damaged or
inoperable. Possible malfunction
of cassette sensor located in
holder.
Summary of Contents for MedSystem III
Page 37: ...34 GETTING STARTED T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K...
Page 45: ...T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K 42 ALARMS ADVISORIES AND PROMPTS...
Page 61: ...58 MAINTENANCE T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K...
Page 63: ...T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K 60 MAINTENANCE...