GETTING STARTED
11
GET
TING ST
ARTED
Prepare solution container in accordance with the
manufacturer’s instructions.
• A syringe can be used as the container for the IV fluid to be
infused. Syringe sizes from 20cc to 60cc of such as the B-D
and Monoject brands can be used.
Connect the container to the IV set.
Prime the MedSystem III® instrument administration set in
accordance with the Administration Set Directions for Use.
It is important to prime the set properly to eliminate air bubbles.
Ensure the cassette slide clamp is pushed in completely so
tubing is not occluded.
Invert the cassette so tubing is up. Slowly open the regulating
clamp and establish fluid flow to fully prime the set. Gently tap
the cassette and ‘Y’ sites as necessary to remove all air. Gently
massage the pressure dome to ensure no air bubbles are
trapped.
1.
Close the regulating clamp before inserting and removing
the cassette to reduce the risk of free flow.
2.
Ensure cassette slide clamp is pulled out (in the closed
position) prior to loading.
3.
Press to turn pump on.
4.
With tubing down, use a 45-degree upward motion to
insert cassette into channel.
5.
Push on clear portion of cassette until completely seated.
Then push in slide clamp flush with entire cassette.
6.
Pull down gently on tubing collar. Press with thumb to seat
tubing collar in recess beneath cassette.
GET
TING ST
ARTED
NOTE
:
The
Model 8631A Syringe Holder
is available as an
accessory that provides a convenient place to hold syringes
while they are being used as containers for IV fluid. The
Syringe Holder is designed to be easily installed and
removed from the top of the pump and to support up to
three syringes. Do not use the Syringe Holder as a handle to
carry the pump.
WARNING
P
Prre
essssu
urre
e D
Do
om
me
e
S
Slliid
de
e C
Clla
am
mp
p
T
Tu
ub
biin
ng
g C
Co
olllla
arr
C
Clle
ea
arr p
po
orrttiio
on
n o
off cca
asssse
etttte
e
NOTE:
Three beeps sound when inserted properly.
A
A
An open regulating clamp and slide
clamp can cause a free-flow
condition and may result in serious
injury to the patient.
Preparing the Infusion
Preparing the Administration Set
Loading the Set
A
Summary of Contents for MedSystem III
Page 37: ...34 GETTING STARTED T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K...
Page 45: ...T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K 42 ALARMS ADVISORIES AND PROMPTS...
Page 61: ...58 MAINTENANCE T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K...
Page 63: ...T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K 60 MAINTENANCE...