MAINTENANCE
45
Check-In
MAINTENANCE
Before unpacking, check the shipping container for damage that
may have affected contents. Report any shipping damage to
Customer Service.
Check to insure that all accessories are included in the package.
Check for any physical damage to the instrument or accessories.
If any is found report it to Customer Service.
Refer to your institutions policies for specific requirements
regarding inspection and testing of incoming equipment before
use. Recommended functional tests are given in the following
pages. As a minimum, the following steps should be performed
before use.
• Charge battery for 14 hours.
• Perform electrical safety checks.
• Turn instrument on to verify normal power-up and
operation of LEDS, display and audio.
This is a Quick Reference Procedure for check-in and
configuration of a new and recently serviced
Medsystem III®
instrument. The following check-in and configuration
procedures are taken from the current service manual.
• Electrical Saftey Test
• Power Tests
• Cassette and Sensor Test
• Patient-side Occlusion Detector Test
• Fluid-side Occlusion Detector Test
• Air-in-Line Test
• Volume Accuracy Test
• Watchdog Audio Test
References (used in conjunction with this document):
• Model 2860 and 2863 Technical Service Manual.
Physical Inspection
Functional Test
Summary of Contents for MedSystem III
Page 37: ...34 GETTING STARTED T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K...
Page 45: ...T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K 42 ALARMS ADVISORIES AND PROMPTS...
Page 61: ...58 MAINTENANCE T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K...
Page 63: ...T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K 60 MAINTENANCE...