ALARMS, ADVISORIES AND PROMPTS
41
ALARMS, ADVISORIES
AND PROMPTS
Fault
Alarm
Meaning
Response
Screen is too light or dark to read
with pump on.
Press .
Press
C
Cn
nt
tr
rs
st
t
softkey to change
screen contrast.
Channel Out of Order
Fault Number
Numeric message, E u r o p e a n s i r e n , r a p i d - flashing red light, infusion stops.
CORRECTIVE ACTION for
resumable faults only.
Press affected channel ,
or .
Follow instructions on channel's
Alarm Information page.
Press
R
RE
ET
TR
RY
Y
to clear fault.
If fault recurs, press
S
SE
ER
RV
VI
IC
CE
E
and
contact qualified service
personnel.
Blank screen, continuous tone red and green lights continuous, all infusions stop.
Watchdog
Other Conditions
Blank Screen
Attempt to reset pump:
Turn pump off, then on again.
Press to resume each
channel that had been infusing.
If Watchdog alarm recurs or
pump cannot be turned on,
replace pump and notify qualified
service personnel.
Pump Shut Off: Low Power.
Pump shut down after a Battery
Depleted alarm had not been
corrected.
Connect AC adapter cord to
pump and plug into wall outlet.
next to External Power
receptacle is lit green when AC
power is properly applied.
Safety checks built into software
have detected a faulty channel.
Safety checks built into software
have detected a fault condition.
Safety checks built into software
have detected an instrument error
condition.
Summary of Contents for MedSystem III
Page 37: ...34 GETTING STARTED T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K...
Page 45: ...T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K 42 ALARMS ADVISORIES AND PROMPTS...
Page 61: ...58 MAINTENANCE T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K...
Page 63: ...T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K 60 MAINTENANCE...